OBJECTIVES The purpose of this study is to investigate dose distribution in the overlap region of two split sub-fields generated by TPS. METHODS IMRT treatment fields larger than 15 cm were planned. TPS created two new sub-fields by splitting each original field. IMRT plans of 14 overlapping complementary binary sub-fields were sent to virtual phantom of TPS for quality assurance (QA) calculations by setting all gantry and collimator angles zero. The measurements at linac were taken using 2D-ion chamber lattice and film both placed at 5 cm depth in solid phantom at QA conditions. RESULTS According to different analyze criterias, consistence of dose fluence related to all fields and sub-fields were found respectively between 81%–99.2% and 80.4%–99.7% by 2D-Array ion chamber system. CONCLUSION As a result there was no statistically significant difference between calculated and measured dose distributions in overlap regions of two split fields derived from fields larger than 15 cm.