In this study, energy resolution dependent on K-Parameters (Trapezoidal filter system optimizing the count rate) of EXOGAM2 detector with newly developed NUMEXO2 digital electronics using 152Eu radioactive source measured and analysed. This experimental study was carried out at the GANIL (Grand Accelerature National d'Ions Lourd) nuclear research center in France. The measurement and analysis results for different K parameters using a 152Eu radioactive source with low and high gamma energy levels were performed for two EXOGAM2 detectors. Without applying any K parameter, the resolution values expressed by FWHM for the 1.EXOGAM2 detector are measured as 2.360% FWHM for (344.2785 keV), 2.820% FWHM for (1408.006 keV), and 2.451% FWHM for (344.2785 keV), 2.975%FWHM for (1408.006 keV) for the 2.EXOGAM2 detector, respectively. For different K parameters (K=2µs, K=5µs ve K=10µs) detail results are shown below.